Feedback solutions for low crosstalk in dense arrays of high-T-c SQUIDs for on-scalp MEG

Silivia Ruffieux (Institutionen för mikroteknologi och nanovetenskap, Kvantkomponentfysik) ; Minshu Xie (Institutionen för mikroteknologi och nanovetenskap, Kvantkomponentfysik) ; Maxim Chukharkin (Institutionen för mikroteknologi och nanovetenskap, Kvantkomponentfysik) ; Christoph Pfeiffer (Institutionen för mikroteknologi och nanovetenskap, Kvantkomponentfysik) ; Alexey Kalabukhov (Institutionen för mikroteknologi och nanovetenskap, Kvantkomponentfysik) ; Dag Winkler (Institutionen för mikroteknologi och nanovetenskap, Kvantkomponentfysik) ; J. F. Schneiderman
Superconductor Science & Technology (0953-2048). Vol. 30 (2017), 5, p. art. nr 054006.
[Artikel, refereegranskad vetenskaplig]

Magnetoencephalography (MEG) systems based on a dense array of high critical temperature (high-T-c) superconducting quantum interference devices (SQUIDs) can theoretically outperform a state-of-the-art MEG system. On the way towards building such a multichannel system, we evaluate feedback methods suitable for use in dense high-T-c SQUID arrays where the sensors are in very close proximity to the head (on-scalp MEG). We test on-chip superconducting coils and direct injection of the feedback current into the SQUID loop as alternatives to the wire-wound copper coils commonly used in single-channel high-T-c SQUID-based MEG systems. For the evaluation, we have performed coupling, noise, and crosstalk measurements. We conclude that direct injection is the optimal solution for dense on-scalp MEG as it gives crosstalk below 0.5% even between SQUIDs whose pickup loops are within 0.8 mm of one another. Further, this solution provides sufficient flux coupling without adding additional noise. Finally, it does not compromise the standoff distance, which is important for on-scalp MEG.

Nyckelord: on-scalp magnetoencephalography, multichannel crosstalk, SQUID, high-temperature



Denna post skapades 2017-06-12. Senast ändrad 2017-09-14.
CPL Pubid: 249692

 

Läs direkt!


Länk till annan sajt (kan kräva inloggning)


Institutioner (Chalmers)

Institutionen för mikroteknologi och nanovetenskap, Kvantkomponentfysik

Ämnesområden

Den kondenserade materiens fysik
Medicinteknik

Chalmers infrastruktur

Relaterade publikationer

Denna publikation ingår i:


Development of high-Tc SQUID magnetometers for on-scalp MEG