Performance of timing resistive plate chambers with relativistic neutrons from 300 to 1500MeV

A. Blanco ; J. Adamczewski-Musch ; K. Boretzky ; P. Cabanelas ; L. Cartegni ; R. F. Marques ; P. Fonte ; J. Fruehauf ; D. Galaviz ; M. Heil ; A. Henriques ; G. Ickert ; D. Korper ; L. Lopes ; M. Palka ; A. Pereira ; D. Rossi ; H. Simon ; P. Teubig ; M. Traxler ; P. Velho ; S. Altstadt ; L. Atar ; T. Aumann ; D. Bemmerer ; C. Caesar ; Alexandre Charpy (Institutionen för fundamental fysik) ; Z. Elekes ; E. Fiori ; I. Gasparic ; J. Gerbig ; K. Gobel ; T. Heftrich ; M. Heine ; Andreas Martin Heinz (Institutionen för fundamental fysik) ; M. Holl ; A. Ignatov ; J. Isaak ; Håkan T. Johansson (Institutionen för fundamental fysik) ; A. Kelic-Heil ; C. Lederer ; Simon Lindberg (Institutionen för fundamental fysik) ; B. Loher ; J. Machado ; J. Marganiec ; Mikael Mårtensson (Institutionen för fundamental fysik) ; Thomas Nilsson (Institutionen för fundamental fysik) ; V. Panin ; S. Paschalis ; M. Petri ; R. Plag ; M. Pohl ; G. Rastrepina ; R. Reifarth ; T. P. Reinhardt ; M. Roder ; D. Savran ; H. Scheit ; P. Schrock ; J. Silva ; D. Stach ; Fredrik Strannerdahl (Institutionen för fundamental fysik) ; Ronja Thies (Institutionen för fundamental fysik) ; A. Wagner ; F. Wamers ; M. Weigand
Journal of Instrumentation (1748-0221). Vol. 10 (2015), p. Art. no. C02034.
[Artikel, refereegranskad vetenskaplig]

A prototype composed of four resistive plate chamber layers has been exposed to quasi-monoenergetic neutrons produced from a deuteron beam of varying energy (300 to 1500 AMeV) in experiment S406 at GSI, Darmstad, Germany. Each layer, with an active area of about 2000x500mm(2), is made of modules containing the active gaps, all in multigap construction. Each gap is defined by 0.3mm nylon mono-filaments positioned between 2.85mm thick float glass electrodes. The modules are operated in avalanche mode with a non-flammable gas mixture composed of 90% C2H2F4 and 10% SF6. The signals are readout by a pick-up electrode formed by 15 copper strips (per layer), spaced at a pitch of 30mm, connected at both sides to timing front end electronics. Measurements of the time of flight jitter of neutrons, in the mentioned energy range, point to a contribution of the resistive plate chamber in the order of 150 ps, independent of the neutron energy.

Nyckelord: Resistive-plate chambers, Instrumentation and methods for time-of-flight (TOF) spectroscopy, Neutron detectors (cold,thermal,fast neutrons); Particle detectors

Denna post skapades 2015-04-22. Senast ändrad 2015-06-23.
CPL Pubid: 215587


Läs direkt!

Länk till annan sajt (kan kräva inloggning)

Institutioner (Chalmers)

Institutionen för fundamental fysik (2005-2015)


Subatomär fysik
Acceleratorfysik och instrumentering

Chalmers infrastruktur