An osseointegrated human-machine gateway for long-term sensory feedback and motor control of artificial limbs

Max Jair Ortiz-Catalan (Institutionen för kliniska vetenskaper, sektionen för anestesi, biomaterial och ortopedi, Avdelningen för ortopedi ; Institutionen för signaler och system, Medicinska signaler och system) ; Bo Håkansson (Institutionen för signaler och system, Medicinska signaler och system) ; Rickard Brånemark
Science Translational Medicine (1946-6234). Vol. 6 (2014), 257, p. Art. no. 257re6.
[Artikel, refereegranskad vetenskaplig]

A major challenge since the invention of implantable devices has been a reliable and long-term stable transcutaneous communication. In the case of prosthetic limbs, existing neuromuscular interfaces have been unable to address this challenge and provide direct and intuitive neural control. Although prosthetic hardware and decoding algorithms are readily available, there is still a lack of appropriate and stable physiological signals for controlling the devices. We developed a percutaneous osseointegrated (bone-anchored) interface that allows for permanent and unlimited bidirectional communication with the human body. With this interface, an artificial limb can be chronically driven by implanted electrodes in the peripheral nerves and muscles of an amputee, outside of controlled environments and during activities of daily living, thus reducing disability and improving quality of life. We demonstrate in one subject, for more than 1 year, that implanted electrodes provide a more precise and reliable control than surface electrodes, regardless of limb position and environmental conditions, and with less effort. Furthermore, long-term stable myoelectric pattern recognition and appropriate sensory feedback elicited via neurostimulation was demonstrated. The opportunity to chronically record and stimulate the neuromuscular system allows for the implementation of intuitive control and naturally perceived sensory feedback, as well as opportunities for the prediction of complex limb motions and better understanding of sensory perception. The permanent bidirectional interface presented here is a critical step toward more natural limb replacement, by combining stable attachment with permanent and reliable human-machine communication.


Video 1: http://youtu.be/Z3uE4bRSkMc

Video 2: http://youtu.be/BgENg8qmzkY



Den här publikationen ingår i följande styrkeområden:

Läs mer om Chalmers styrkeområden  

Denna post skapades 2014-10-17. Senast ändrad 2015-07-01.
CPL Pubid: 204414

 

Läs direkt!


Länk till annan sajt (kan kräva inloggning)


Institutioner (Chalmers)

Institutionen för kliniska vetenskaper, sektionen för anestesi, biomaterial och ortopedi, Avdelningen för ortopedi (GU)
Institutionen för signaler och system, Medicinska signaler och system

Ämnesområden

Livsvetenskaper
Materialvetenskap
Medicinteknik
Ortopedi

Chalmers infrastruktur

Relaterade publikationer

Denna publikation ingår i:


Towards Natural Control of Artificial Limbs