Feedback Analysis in Percutaneous Bone-Conduction Device and Bone-Conduction Implant on a Dry Cranium

Hamidreza Taghavi (Institutionen för signaler och system, Medicinska signaler och system) ; Bo Håkansson (Institutionen för signaler och system, Medicinska signaler och system) ; Sabine Reinfeldt (Institutionen för signaler och system, Medicinska signaler och system) ; Måns Eeg-Olofsson ; Shirin Sadat Akhshijan (Institutionen för signaler och system, Medicinska signaler och system)
Otology and Neurotology (1531-7129). Vol. 33 (2012), 3, p. 413–420.
[Artikel, refereegranskad vetenskaplig]

Hypothesis: The bone-conduction implant (BCI) can use a higher gain setting without having feedback problems compared with a percutaneous bone-conduction device (PBCD). Background: The conventional PBCD, today, is a common treatment for patients with conductive hearing loss and single-sided deafness. However, there are minor drawbacks reported related to the percutaneous implant and specifically poor high-frequency gain. The BCI system is designed as an alternative to the percutaneous system because it leaves the skin intact and is less prone to fall into feedback oscillations, thus allowing more high-frequency gain. Methods: Loop gains of the Baha Classic 300 and the BCI were measured in the frequency range of 100 to 10,000 Hz attached to a Skull simulator and a dry cranium. The Baha and the BCI positions were investigated. The devices were adjusted to full-on gain. Results: It was found that the gain headroom using the BCI was generally 0 to 10 dB better at higher frequencies than using the Baha for a given mechanical output. More specifically, if the mechanical output of the devices were normalized at the cochlear level the improvement in gain headroom with the BCI versus the Baha were in the range of 10 to 30 dB. Conclusion: Using a BCI, significantly higher gain setting can be used without feedback problems as compared with using a PBCD.

Nyckelord: Baha, Bone-conduction implant, Feedback, Gain headroom, Implantable transducer, Loop gain, Percutaneous bone-conduction device, Sound radiation, Stability



Denna post skapades 2012-03-13. Senast ändrad 2016-06-30.
CPL Pubid: 155866

 

Läs direkt!

Lokal fulltext (fritt tillgänglig)

Länk till annan sajt (kan kräva inloggning)


Institutioner (Chalmers)

Institutionen för signaler och system, Medicinska signaler och system
Institutionen för kliniska vetenskaper, sektionen för anestesi, biomaterial och ortopedi, Avdelningen för öron-, näs- och halssjukdomar (GU)

Ämnesområden

Elektronisk mät- och apparatteknik

Chalmers infrastruktur