Electromagnetic Modeling and Design of Medical Implants and Devices

Oskar Talcoth (Institutionen för signaler och system, Biomedicinsk elektromagnetik)
Göteborg : Chalmers University of Technology, 2011. - 147 s.

This thesis covers two topics in biomedical electromagnetics: pacemaker lead heating in magnetic resonance imaging (MRI) and optimization of sensor positions in magnetic tracking.

The electromagnetic part of pacemaker lead heating during MRI is a resonant phenomenon which is complicated by, among other factors, the wide range of length scales involved in the problem. In this work, the multi-scale part of the problem is taken into special consideration during the modeling process. The model incorporates a radio frequency coil, a human body phantom,and a highly detailed model of a pacemaker system with a bipolar lead that features helix-shaped conductors.

Several configurations of pacemaker systems exposed to MRI are modeled and the results clearly show the importance of detailed lead modeling. Furthermore, modeling of resonant structures is investigated by a comparison between different modeling techniques. In addition, a meshing scheme for thin-wire approximations of helices is proposed, evaluated, and found to have improved convergence properties as compared to the conventional meshing approach.

In recent years, magnetic tracking has been applied in many biomedical settings due to the transparency of the human body to low-frequency magnetic fields. In this work, the sensor positions of a magnetic tracking system are optimized by exploiting an analytical model where the transmitting and sensing coils of the system are approximated by magnetic dipoles.

In order to compare different sensor array layouts, two performance measures based on the Fisher information matrix are discussed and compared for the optimization of the sensor positions of a circular sensor array. Furthermore, the sensor positioning problem is formulated as an optimization problem which is cast as a sensor selection problem. The sensor selection problem is solved for a planar sensor array by the application of a convex relaxation. Several transmitter positions are considered and general results are established for the dependence of the optimal sensor positions on the transmitter’s position and orientation.

Nyckelord: Helix, Implant Safety, Lead Heating, Magnetic Resonance Imaging, MR Safety, Pacemakers, Thin-wire Approximation, Magnetic Tracking, Sensor Selection, Fisher Information Matrix, Convex Relaxation, Optimal Sensor Placement, Optimal Measurements

Den här publikationen ingår i följande styrkeområden:

Läs mer om Chalmers styrkeområden  

Denna post skapades 2011-12-05. Senast ändrad 2011-12-06.
CPL Pubid: 149505


Läs direkt!

Lokal fulltext (fritt tillgänglig)

Institutioner (Chalmers)

Institutionen för signaler och system, Biomedicinsk elektromagnetik


Elektroteknik och elektronik

Chalmers infrastruktur

Relaterade publikationer

Inkluderade delarbeten:

Monolithic Multi-Scale Modeling of MR-Induced Pacemaker Lead Heating

Electromagnetic Modeling of Pacemaker Lead Heating During MRI

Optimization of Sensor Positions in Magnetic Tracking


Datum: 2011-12-19
Tid: 13:00
Lokal: Room EC, EDIT building, Hörsalsvägen 11, Chalmers University of Technology
Opponent: Prof. Mats Gustafsson, Department of Electrical and Information Technology, Lund University, Lund, Sweden

Ingår i serie

R - Department of Signals and Systems, Chalmers University of Technology