Experimental Validation of a Method for Removing the Capacitive Leakage Artifact from Electrical Bioimpedance Spectroscopy Measurements

R. Buendia ; Fernando Seoane (Institutionen för signaler och system, Medicinska signaler och system) ; Roberto gil-pita
Measurement science and technology (0957-0233). Vol. 21 (2010), 11, p. 115802.
[Artikel, refereegranskad vetenskaplig]

Often when performing electrical bioimpedance (EBI) spectroscopy measurements, the obtained EBI data present a hook-like deviation, which is most noticeable at high frequencies in the impedance plane. The deviation is due to a capacitive leakage effect caused by the presence of stray capacitances. In addition to the data deviation being remarkably noticeable at high frequencies in the phase and the reactance spectra, the measured EBI is also altered in the resistance and the modulus. If this EBI data deviation is not properly removed, it interferes with subsequent data analysis processes, especially with Cole model-based analyses. In other words, to perform any accurate analysis of the EBI spectroscopy data, the hook deviation must be properly removed. Td compensation is a method used to compensate the hook deviation present in EBI data; it consists of multiplying the obtained spectrum, Zmeas(ω), by a complex exponential in the form of exp(–jωTd). Although the method is well known and accepted, Td compensation cannot entirely correct the hook-like deviation; moreover, it lacks solid scientific grounds. In this work, the Td compensation method is revisited, and it is shown that it should not be used to correct the effect of a capacitive leakage; furthermore, a more developed approach for correcting the hook deviation caused by the capacitive leakage is proposed. The method includes a novel correcting expression and a process for selecting the proper values of expressions that are complex and frequency dependent. The correctness of the novel method is validated with the experimental data obtained from measurements from three different EBI applications. The obtained results confirm the sufficiency and feasibility of the correcting method.

Nyckelord: electrical bioimpedance spectroscopy, capacitive leakage, artifact removal



Den här publikationen ingår i följande styrkeområden:

Läs mer om Chalmers styrkeområden  

Denna post skapades 2011-01-24. Senast ändrad 2017-08-18.
CPL Pubid: 136030

 

Läs direkt!


Länk till annan sajt (kan kräva inloggning)


Institutioner (Chalmers)

Institutionen för signaler och system, Medicinska signaler och system (2005-2017)

Ämnesområden

Informations- och kommunikationsteknik
Elektronisk mät- och apparatteknik
Medicinsk teknik

Chalmers infrastruktur